Sapphire 800 Analyser
The Glenbio SAPPHIRE 800 Auto-Chemistry Analyser is an instrument with discrete system,
reagent open function, emergency priority function as well as an external computer.
The instrument is composed of humanized software operation system, intelligent zed optical unit, complicated mechanism system, precision liquid path and accuracy electrical system.
The instrument could automatically realize sampling, reagent injection, anti-interference, mixture, pre-temperature, reaction measurement, rinse, calculation, display and print function. The substitution of manual operation for automatic operation could not only enhance the working efficient but alsodecrease the test error, thus greatly enhance the accuracy and precision of test results.
SAPPHIRE 800 Auto-Chemistry Analyser could carry out the immunology check and biochemical analyze of blood, urine, ascites, cerebrospinal fluid and other body fluid. The instrument could also
carry out clinic test, such as: myocardium enzymogram, blood sugar, blood fat, liver function, renal
function, immunoglobulin, etc.
TECHNICAL SPECIFICATIONS
-
- Instrument structure: Discrete system
- Throughput: Constant speed, 400 tests/ hour (800 tests/ hour with ISE)
- Simultaneous analysis item: At most 88 colorimetric items, 3 ISE items (K, Na, Cl).
- Sample volume: 2 to 35µl (Stepping 0.1µl)
- Reagent volume: 20 to 350µl (Stepping 1µl)
- Reaction solution volume: 150~450µl
- Liquid level sensor: Integration of sample probe and reagent probe with touch
- sensor and sample probe block test function.
- Stirring: Independent stirring after reagent injecting
- Sample disk: 115 samples position (50 routine samples, 34 Calibrators, 20
- STAS samples, 8 Controls, 3 Detergents)
- Reagent disk: Dual disk
- Disk 1: 45 positions for Reagent 1, Reagent 4, diluents and CS antibacterial phosphor-free detergent
- Disk 2: 45 positions for Reagent 2, Reagent 3, CS anti-bacterial phosphor-free detergent Photometer: Grating spectrophotometry system in a range of 340~750nm, wavelength: 340, 380, 405, 450, 480, 505, 546, 570, 600, 660, 700, 750nm
- Wave length accuracy: ±2nm
- Light source: 20W /12V Long life quartz halogen lamp (water cooling)
- Measurement range: 0 to 3.3Abs
- Reaction disk: 120 pcs of reusable rigid optical plastic reaction cuvette.
- Reaction cuvette optical diameter: 6mm
- Reaction cuvette rinse: Automatic
- Incubation bath temperature: 37°±0.1°
- Reaction time: 15 minutes at most (optional for 3, 4, 5, 10 and 15 minutes)
- Analysis method: Rate assay, end-point assay, 2-point assay.
- Calibration method: 1-point linearity, 2-point linearity, multi-point linearity, nonlinearity method.
- Reagent bottle volume: 20ml, 70ml
- Reagent cooling unit: All reagents keep at 5°- 15°, semiconductor refrigeration.
- Barcode scanning: 3 internal barcode scanner (scan the barcode on the routine sample, and reagent on the R1, R2 disk)
- Reagent volume test: Test and report the reagent remaining volume.
- Power supply: ~220/230V 50Hz
- Ambient temperature: 15°~32° Optimum temperature: 18° to 25°
- Relative humidity: 40% ~ 85%
- Appearance dimension: 1060×790×1150mm (length×width×height)
- Fixing power: 2000VA
- Weight: About 300Kg