Beckman Coulter DxC 700 AU
The DxC 700 AU delivers increased uptime, high reliability and precise performance to mid- to large-sized laboratories and hospitals. Its flexible design offers standalone operation or connectivity to laboratory automation systems. With throughput of up to 800 photometric tests per hour (up to 1,200 with ISEs) and 63 onboard parameters, the DxC 700 AU delivers reliability and efficiency to laboratories around the world.
INSTRUMENT SPECIFICATIONS
- Newly designed intuitive graphical user interface, including simplified reagent inventory, calibration and quality control procedures
- Online maintenance documentation
- User-customized menu and free-text notepad
- Colour alerts to highlight system operating conditions
- Reliability for greater uptime with quick and easy replacement of common parts
- No special tools required
- No more than three steps and no longer than 60 seconds for replacement of parts, such as sample and reagent probes, mixers, syringes, ISE electrodes and optical lamp
- Extensive online help with instructional videos

- Available integration with pre-analytical automation through direct track sampling from Power Express
- Available connection to Command Central with REMISOL Advance, allowing access to multiple instruments and automation consoles from a single workstation, optimizing laboratory management and improving decision making
- PROService remote diagnostic capabilities ensure 24/7 continuous monitoring of vital system functions
- Cooled STAT compartment, providing one-button STAT interrupt, advanced auto QC and calibration capabilities
- Ability to load reagents while the analyzer is in measure mode
- Priority sample repeat/reflex
- Economical ISEs with long onboard stability; easy to maintain (only individual electrode replacement required)
- 150-sample continuous rack loader
MORE SPECIFICATIONS
Analytical system
Fully automated, random-access clinical chemistry system with
STAT capability
Analytical principles
Spectrophotometry and potentiometry
Assay types
Endpoint, rate, fixed point and indirect ISE
Analytical methods
Colorimetry, turbidimetry, latex agglutination and indirect ISE
Test menu applications
I20
Programmable tests
117 photometrics (113 + LIH + HbA1c (Thb, HbA1c + HbA1c%)) and ISE
Onboard parameters
60 photometric tests and 3 ISEs (Na, K, Cl)
Throughput
800 photometric tests per hour, up to 1,200 with ISE
ISE sample throughput: 200 per hour
ISE maximum tests per hour: 600 if ISE only
Sample types
Serum, plasma, urine, whole blood (HbA1c) and other fluids
Sampler capacity
Rack sampler: 10 samples per rack (barcodes on primary tubes and
on racks), capacity for 150 samples
Refrigerated STAT carousel: 22 samples can be run simultaneously
(cal, QC and routine samples)
Sample tubes
Primary and secondary tubes: diameter between 11.5 mm and
16 mm; height between 55 mm and 102 mm; nested micro cups
Sample volume
- μL-25 μL in 0.1 μL increments
Sample quality analysis
Lipemia, hemolysis and icterus indices
Clot detection and probe crash protection
Sample bar-code formats
NW7, EAN13, CODE 39, CODE 128, ISBT-128, 2 of 5 standard, 2 of 5
interleaved mixed readable (max. four types at the same time, except
if using ISBT-128)
Reagent supply
60 positions for R1, 48 positions for R2 (refrigerated 4°C-12°C); bottle
sizes: 15 mL, 30 mL, 60 mL
Reagent volume
R1: 10 μL-250 μL; R2: 10 μL-250 μL (1 μL increments)
Total reaction volume
120-350 μL
Reaction cuvette
Permanent glass cuvettes
Reaction time
Up to 8 minutes, 33 seconds
Reaction temperature
37°C ±0.3°C
Reaction method
Dry bath
Photometric range
0 OD-3.0 OD
Wavelength
13 different wavelengths between 340 nm and 800 nm
Calibration
Auto-calibration, advanced calibration and cooled calibrator
positions
Calibrator value assignment by 2D barcode; 200 calibrators can
be programed
History of graphical calibration data stored
Quality control
Westgard Rules, Twin Plot and Levey-Jennings graphs, auto QC,
cooled QC positions
100 controls can be programed, 10 levels per test
Reflex testing
User-defined
Automated sample pre-dilution
Repeat run with increased or decreased sample volume or sample
pre-dilution (3, 5, 10, 15, 20, 25, 50, 75, up to 100 times)
Online
Uni- and bi-directional host query communications
Operating system
Windows 7
Data storage
Up to 100,000 patient samples; reaction monitor: 400,000 tests,
300 indexes
INSTALLATION REQUIREMENTS
Dimensions (W x H x D)
Analyzer:
49 in x 51 in x 35 in (1250 mm x 1300 mm x 890 mm)
1,015 lbs (460 kg)
Sampler:
29 in x 37 in x 41 in (730 mm x 940 mm x 1040 mm)
331 lbs (150 kg)
Power supply
200V, 208V, 220V, 230V, 240V, 50 Hz, 60 Hz, 3.8 kVA
Water supply information
Mean water consumption: 28 L per hour
Water type: deionized CAP Type II, bacteria free
Continuous flow supply
Resistivity: less than 2.0 μS/cm filtered with a 0.5 μm filter
Temperature and humidity
18°C to 32°C, 20% to 80% RH (no condensation)
Drain requirements
Built-in waste pump
Drain required: maximum height from floor < 59 in (< 1.5 m)